Femoral Access, Insertion
Impella RP Flex™ with SmartAssist™ Insertion Animation
This animation depicts the primary steps for femoral insertion of the Impella RP FlexTM with SmartAssistTM. Please refer to the instructions for use manual for a full list of steps. The Impella RP Flex with SmartAssist is designed for right-heart recovery and is indicated for use for up to 14 days. For more information, consult the Impella RP Flex with SmartAssist Instructions for Use manual at www.heartrecovery.com.
1. Gain access to the femoral vein.
- Using ultrasound, gain access to the femoral vein with a micropuncture needle and sheath kit. Exchange for a 5-8 Fr introducer sheath. Insert a 0.035” guidewire.
Place a hemostatic stitch around the sheath but do not secure.
2. Prepare the vasculature for insertion of the Impella RP Flex with SmartAssist.
- Flush the sheath per standard hospital practice and sequentially dilate the puncture using the 8, 12, 16, 20, and 23 Fr dilators in the Introducer kit.
- Insert the 23 Fr peel-away sheath. Ensure ACT >250 seconds. Remove the dilator and guidewire. Attach a flush syringe to the 23Fr introducer sidearm and aspirate and flush the 23 Fr introducer.
- Under fluoroscopic guidance advance a compatible flow directed catheter into the femoral vein and navigate to the right or left pulmonary artery.
- Create a small “J” on the 0.027” Impella guidewire and advance into the pulmonary artery.
- Remove the flow directed catheter using a fixed-wire technique.
3. Insert the Impella RP Flex with SmartAssist.
- Backload the Impella RP Flex cannula onto the 0.027” guidewire through the red EasyGuide lumen. Once backloaded, remove the red EasyGuide lumen.
- Using fixed-wire technique under fluoroscopic guidance, advance the Impella RP Flex with SmartAssist through the sheath and into the inferior vena cava (IVC). Continue to advance the Impella RP Flex until the inlet is in the IVC and the outlet is approximately 4 cm past the pulmonary valve annulus.
- Slowly remove the guidewire under fluoroscopic guidance. Confirm position with fluoro.
4. Start support with the Impella RP Flex with SmartAssist.
- Start the Impella RP Flex at performance level (P-level) P-2.
- Slowly increase support to the desired P-level.
5. Secure the position of the Impella RP Flex with SmartAssist.
- Remove the sheath from the femoral vein and peel-away. Use fluoroscopy to ensure that the Impella outlet does not move.
Grasp the two wings and bend back until the valve assembly splits. To do this, first stretch then snap the flexible valve mechanism that temporarily holds the two wings together. Continue to peel the two wings until the introducer is completely separated from the catheter shaft.
Note: Do NOT peel the 23 Fr peel-away introducer over the tip of the repositioning sheath.
Place two dead-end caps on the repositioning sheath stopcock to prevent further usage. The sideport should not be used to give medication or draw blood because the blood could potentially clot. Pressure bags should not be connected to the sideport of the repositioning sheath.
- Slide the repositioning sheath into the vein under fluoroscopy, ensuring the Impella cannula does not move. Tighten the hemostatic stitch. Advance the sterile sleeve and turn it clockwise to lock it to the suture hub.
- Confirm all slack has been removed. Tighten the Tuohy Borst valve by turning clockwise until finger-tight.
- If needed, maintain the suture hub angle with folded 4x4 gauze before placing the sterile dressing.
- Secure the Impella RP Flex with SmartAssist per hospital protocol.
Note: Do NOT clean the Impella handle or purge sidearm retainer with alcohol or alcohol-based products.
IMP-6813